Mã SP:
P2VA1
- Model:
Valcom
Cảm biến đo áp suất P2VA1 Valcom
|
Measured medium |
Water, oil, gases or other media that will not corrode SUS. |
Pressure type |
Positive pressure can be measured. |
Measured Range |
Rated capacity: 2000, 3000, 5000bar |
Output type |
DC0.5~10V |
Specification(Mechanical input characteristics)
Pressure type |
Absolute pressure |
Accuracy Class |
0.3%R.C. |
Initial Value |
0bar |
Operating Range at Reference Temperature |
0 ~ Approx. 110% |
Overload Limit at Reference Temperature |
150% |
Test Pressure |
200% (5000bar: 150%) |
During Dynamic Loading |
Permissible Pressure |
100% |
Permissible Vibration Amplitude
(As per DIN50, 100) |
70% |
Dead Volume |
0.8cm³ |
Control Volume |
1.5mm³ |
Materials of Measurement Media's Contact Part |
1.4542(Measurement body) 1.4301(Seal) |
- * This is a made-to-order model. Contact our sales office for details.
Specification(Output characteristics)
Transducer Identification |
TEDS (*1) |
Signal Span(Sensitivity) |
0.5~10V |
Sensitivity Tolerance of Zero Signal (Factory) |
2000, 3000bar: <±0.010V
5000bar: ±0.020V |
Sensitivity Tolerance |
2000, 3000bar: <±0.010V
5000bar: ±0.020V |
Maximum Signal |
10.5V |
Temperature Effect on Zero Balance (*2) |
0.2%/10k |
Temperature Effect on Load (*2) |
0.2%/10k |
(Start Setting)Characteristic Curve Deviation |
0.3% |
Repeatability (As per DIN1319) |
<±0.05% |
Cut-off Frequency of Output Signal |
-3dB |
4.5kHz |
-1dB |
2kHz |
Burden (Minimum) |
10000Ω |
|