Máy đo chiều dày lớp sơn, mạ CT-3 Densoku


Mã SP: - Model: Densoku

Máy đo chiều dày lớp sơn, mạ CT-3 Densoku


Yêu Cầu Giá Tốt

    • SĐT: 0948.007.822
    • Mail: yen@ttech.vn
      • Zalo: 0948.007.822
THÔNG TIN SẢN PHẨM

Electrolytic film thickness meter CT-3 Features

Lightweight, compact

Gunto lightweight, compact body.

Corresponding to various plating measurement

It can be measured even from any plating range.

High sensitivity range added

It adds a high sensitivity range.

A wide range of cab rate range

Cab rate range is extensive and 15% ±.

1/100 range is possible

Au, Cr Similarly, other plating is also possible to 1/100 range.

C gasket range standard equipment

A, to B gasket, C gasket range is also standard equipment.

Active featured

With an active function, also measures to smooth sample was somewhat oxidation.

Easy range setting

Easy range setting of measurement plating, is a mistake hard to conscience design.

Electrolytic film thickness meter CT-3 accessories

accessories
Measuring table Settings are very easy! Round bar is also at the touch of a button OK! In particular, it is ideal for continuous measurement of the same shape.
Standard accessory - Electrolyte solution (100ml) 1 present 
and washing bottle 1 pc 
and liquid bottle 1 pc 
, measuring table 
cell A, B 
-gasket A, B, C 
, standard plate (Ni / Fe) 1 sheets, cleaning fluid (100ml) 1 this, beaker dropper, stirring tube equation (1), fuse, manual

Option of electrolytic film thickness meter CT-3

option
WT The following of the wire and the width 1.7mm, for the measurement of small parts requires WT (wire tester).
Standard accessories 
(optional)
· Positive cathode jig · WT (wire tester)


standard

Film thickness meter of Densoku is, JIS (Japan standard industrial) H8501, H8610-8619, STM (American Society for Testing and Materials) B504-82, MIL (US military standard), ISO (International Standard) 2177, DIN (German standard) conform to standards such as 50932,50955, it is an international measuring instrument that has been certified with the US Kokuru Corporation.

Specification of electrolytic film thickness meter CT-3

Measurement range 0.006 ~ 300μm
Minimum resolution 0.125μ / sec, or 0.0125μ / sec, 0.00125μ / sec.
Body accuracy ± 1%
Unit of measurement μm, nm (direct reading by the LED display)
Measurement area 3.4mmφ If you use the A-type gasket 
2.4mmφ when using a B-type gasket 
if you use the C-type gasket 1.7mmφ
Power supply AC100V (115V, 200V, 230V), 50 / 60Hz
Body dimensions H121 × W240 × D181mm
weight 3.0kg (body only)

(Note) The specifications are subject to change without notice.

Various combinations and electrolytic reagent of the green body and the plating

Kind of plating Electrolyte Green body
copper K-44 Iron, aluminum, nickel, permalloy, silver, Kovar, molybdenum, stainless steel, 
non-metallic
K-52 Brass, beryllium copper, lead, zinc die-cast, nickel silver, tin, cadmium
nickel K-54 Iron, aluminum, brass, copper, inconel, molybdenum, tungsten, stainless steel, 
non-metallic, chrome
Electroless nickel K-57 Iron, aluminum, Kovar, non-metal, stainless steel, molybdenum
Chrome K-51 Iron, aluminum, nickel, stainless steel, non-metallic, Inconel
zinc K-46 Iron, aluminum, brass, copper, nickel, non-metallic, tin, tungsten
tin K-47G Iron, nickel, nickel silver, nickel silver, permalloy, cadmium, non-metallic
K-51 Aluminum
K-50 Copper, brass
Silver K-44 Iron, aluminum, non-metal, permalloy, Kovar
K-48 Brass, copper, tin, non-metallic
K-48G Nickel, nickel silver, nickel silver
solder K-44 Iron, aluminum, brass, copper, metallic
cobalt K-54 Iron, brass, copper, metallic
brass K-44 Iron, aluminum, non-metallic
Surface treatment solution N-64 Deposit removal solution after the nickel measurement
CS For polishing of the dirt off and the resin of the cleaning suspension ... the object to be measured